8.6 Message Reference
146
Continuous Measurement Setting
Syntax
Command
:INITiate:CONTinuous <1, 0, ON or OFF>
Query
:INITiate:CONTinuous?
Response <ON or OFF>
ON......... Continuous Measurement Enabled
OFF....... Continuous Measurement Disabled
Description
Command Sets continuous measurement.
Query Queries the continuous measurement setting.
Example
Command
:INIT:CONT OFF
Disables continuous measurement.
Query
Response
:INIT:CONT?
ON
Enables continuous measurement.
Note
Continuous Measurement Enabled:
After measurement, enters the Trigger Wait State. When the trigger source set-
ting is IMMediate, the next trigger occurs immediately (the Free-Run State).
Continuous Measurement Disabled:
After measurement, enters the Idle State instead of the Trigger Wait State.
Triggering is ignored in the Idle State. Executing :INITiate[:IMMediate] enables
the Trigger Wait State.
Continuous measurement is enabled upon exit from the Remote State.
Trigger Wait Setting
Syntax
Command
:INITiate[:IMMediate]
Description
Command Switches triggering from the Idle State to the Trigger Wait State.
Example
Command Disable continuous measurement, and read one value for each trigger
event
Send
:TRIG:SOUR IMM....Trigger immediately when entering Trigger
Wait State
:INIT:CONT OFF....Disables continuous measurement
:INIT ...........................Enable Trigger Wait Trigger immediately
upon :TRIG:SOUR IMM
:FETC? ........................Fetch measured value
Response
2.1641E+0.................Measured value is 2.1641 Ω
Error
An execution error occurs when continuous measurement is enabled (:INI-
TIATE:CONTINUOUS ON).
Note
When the trigger source is IMMediate, triggering occurs immediately before
entering the Idle State.
When the trigger source is EXTernal, the Trigger Wait State is enabled to wait
for an external trigger, and when a trigger occurs, one measurement is taken
before entering the Idle State.