Loading...
113
Data Output Settings
9.3 Data Output Settings
(Measurement screen) MENU > I/F
1
Enable automatic output.
(Default setting:
OFF)
When automatic output is set to ON, do not
use communications commands. Doing so
may cause measured value data to be sent
twice.
2
(To change the information that will be
output)
1
2
MEASUREMENT
ITEM
V: Voltage value (default
setting)
V,TEMP
: Voltage value
and temperature
DATE-TIME Measurement time and
date (default setting: OFF
[output disabled])
COMP
CONDITION
ALL: All judgments (default
setting)
HI: HI judgments
IN: IN judgments
LO: LO judgments
HL: HI and LO judgments
If comparator judgment or BIN measurement
is set to ON, judgment results will also be
output.
When the interface is set to USBKEYBD,
DATE-TIME is not output.
3
(To change the output format)
See “7.8 Setting Output Formats” (p. 92).