167
14
Maintenance and Service
WARNING
Customersshouldnotattempttomodify,disassembleorrepairtheinstrument.
Fire,electricshockandinjurycouldresult.
Calibration and repair
The calibration frequency varies depending on the status of the instrument or installation
environment. We recommend that the calibration frequency is determined in accordance with
the status of the instrument or installation environment and that you request that calibration be
performed periodically.
When calibration or repair is requested to Hioki, the settings will be returned to the default settings.
Before requesting calibration or repair, we recommend saving the settings of the instrument in a
USB ash drive.
Transporting the instrument
To ensure that the product arrives safely, use the original box and packaging from when you
purchased it. However, do not use the original box if it is torn or otherwise damaged or the original
packaging if it has been crushed. Instead, use standard, commercially available packaging
materials to carefully pack the product in the same manner as it arrived after purchase.
Please note that if you pack the product so that it is not adequately cushioned and it suffers
damage during shipment, you will be billed for the cost of repair, even if the product is still within
the warranty period.
Be sure to disconnect all cables from the product before packing it.
Exercise care that the product is not dropped or subject to other mechanical shock during
shipment.
Replacement parts and their service lives
The characteristics of some of the parts used in the product may deteriorate with extended use.
To ensure the product can be used over the long term, it is recommended to replace these parts on
a periodic basis.
When replacing parts, please contact your authorized Hioki distributor or reseller.
The service life of parts varies with the operating environment and frequency of use. Parts are not
guaranteed to operate throughout the recommended replacement cycle.
Part Name
Recommended
ReplacementCycle
Notes and Conditions
Electrolytic
capacitors
Approx. 5 years The PCB on which the part is mounted must be replaced.
LCD backlight (half-
life of brightness)
Approx. 5 years Based on 24 hours/day usage
Relays Approx. 5 years For range switching 10 times/h
Backup battery
(lithium battery)
Approx. 10 years If the date or time is not substantially accurate, the battery
should be replaced.