126
Switching between Current Sink (NPN) and Current Source (PNP)
11.2
Switching between CurrentSink(NPN) and
Current Source (PNP)
Before using, read “Before switching the current sink (NPN) / current source (PNP)” (p. 11)
carefully.
The NPN/PNP switch is used to change the type of programmable controller that can be supported.
The instrument ships with the switch in the NPN position.
Left: Current sink (NPN)
Right: Current source (PNP)
See “Internal circuit architecture” (p. 133).
NPN/PNPswitchsetting
NPN PNP
Input circuit
Supports programmable controllers
that generate sink output.
Supports programmable controllers
that generate source output.
Output circuit Non-polar Non-polar
ISO_5V power supply output +5 V output -5 V output
w ww . . co m