52
To Obtain Accurate Judgments Even When the Polarity of the Measurement Target (Battery, etc.) Is Reversed
4.1 To Obtain Accurate Judgments Even When the
PolarityoftheMeasurementTarget(Battery,
etc.) Is Reversed
By setting the absolute value judgment function to ON, voltage can be judged as a positive value
even when it is negative.
(Measurement screen) MENU > MATH > ABS MODE
(Default setting:
OFF)
Whentheabsolutevaluejudgmentfunctionisenabled(examplewhenboththeupperlimit
and lower limit values are positive)
Upper limit and lower limit values
Upper limit
HI
IN
Lower limit
LO
0
- Lower limit
LO
IN
- Upper limit
HI
When (lower limit value ≤ |measured
value| ≤ upper limit value), the
instrument will generate an IN
judgment.
Upperlimitvalueonly
Upper limit
HI
IN
0
- Upper limit
IN
HI
When (|measured value| ≤ upper
limit value), the instrument will
generate an IN judgment.
Lowerlimitvalueonly
Lower limit
IN
LO
0
- Lower limit
LO
IN
When (|measured value| ≥ lower limit
value), the instrument will generate
an IN judgment, including in the
event of an over-range (+OvrRng/-
OvrRng) event.
w ww . . co m