61
5
Saving and LoadingMeasurement
Conditions(InternalMemory)
The current measurement conditions can be saved to the instrument’s internal memory (via the
panel save function) and loaded from the instrument’s internal memory (via the panel load function)
as follows:
• By means of touch panel operation
• By sending communications commands from an external device
• By sending signals from an external device
A maximum of 30 panels (with panel numbers 01 through 30) can be saved. Panel data is
maintained even when the instrument is turned off.
Measurement conditions can also be saved on a USB ash drive. See “10 Outputting and
Loading Measurement Conditions (USB Flash Drive)” (p. 120).
Information that can be saved with the panel save function
Save time and date Measured value
display
Range selection Input resistance
selection
Numberofdisplay
digits
Integration time Smoothing Trigger settings
(measurement count
anddelay)
NULL Temperature
compensation
Scaling Contactcheck
Comparator BIN Absolute value
judgment
Auto-hold
Labeldisplay Sub-display
5 Saving and Loading Measurement Conditions (Internal Memory)
5
Saving and Loading Measurement Conditions (Internal Memory)