EasyManua.ls Logo

Hioki DM7275-01 - X84; Accuracy Specifications

Hioki DM7275-01
207 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
155
Measurement Specications
Accuracyspecications
Conditions of guaranteed
accuracy
Guaranteedaccuracyperiod:
1 year
Temperatureandhumidityforguaranteedaccuracy:
23°C ±5°C (73°F ±9°F), 80% RH or less
Warm-up time: 1 hr.
Measurement cable
Low thermal electromotive force cable (FLUKE 5440A-7005)
Voltage measurement
accuracy
DM7275-01,DM7275-02,DM7275-03:
See “Separate table 1 (DM7275)” (p. 156).
DM7276-01,DM7276-02,DM7276-03:
See “Separate table 2 (DM7276)” (p. 156).
Additional errors:
Temperature coefcient
From 0°C to 18°C and from 28°C to 40°C, add the following value per 1°C of
temperature:
100 mV to 10 V range:
±
0.05
×
measurement accuracy/°C
100 V, 1000 V range:
±
0.1
×
measurement accuracy/°C
Voltage coefcient error
Add the following value to the rdg. error component if the voltage display value
Vin exceeds ±300 V:
DM7275: 0.0010% × (Vin / 1000)
2
DM7276: 0.0005% × (Vin / 1000)
2
Noise error (excluding effects of burst noise)
Integration time
T
I
Additional error
10 PLC ≤
T
I
None
1 PLC ≤
T
I
< 10 PLC 0.0001% of range ± 0.5 µV
0.2 PLC ≤
T
I
< 1 PLC 0.0003% of range ± 1 µV
0.02 PLC ≤
T
I
< 0.2 PLC 0.0010% of range ± 2 µV
Temperature compensation error
When using temperature compensation, add the following value to the
resistance measurement accuracy’s rdg. error component:
( )
0
100[%]
1
T
T T T
α
α
×
+ × +
T
0
: Reference temperature [°C]
T
: Current ambient temperature [°C]
T
: Temperature measurement accuracy
α
: Temperature coefcient [1/°C]
Measurement cable error
Temperature difference of no greater than 1°C between the instrument and the
measurement cable/measurement target
Add individual error components if connecting cables in series.
L9207-10 Test Lead
L4933 Contact Pin Set
L4932 Test Pin Set
10 µV
L4934 Small Alligator Clip Set
L4935 Alligator Clip Set
7 µV
9243 Grabber Clip
L4936 Bus Bar Clip Set
5 µV
L4931 Extension Cable Set 3 µV
L4930 Connection Cable Set 2 µV
Effects of radiative radio-frequency magnetic eld: 3% of range at 10 V/m
Effects of conductive radio-frequency magnetic eld: 3% of range at 3 V
13
Specications
w ww . . co m
information@itm.com1.800.561.8187