Measurement Specications
Parameter
Description Time
T
6
Acquision time (RUN state)
50 Hz power 60 Hz power
FAST (1PLC) 26.9 ms 23.5 ms
MEDIUM (10PLC) 245 ms 205 ms
SLOW (100PLC) 3.92 s 3.37 s
Integration time other than above: Integration time + 5 ms
T
7
Calculation time 0.1 ms
T
8
From EOM signal output to TRIG signal
input
1 ms or greater
T
9
EOM pulse width (other than RUN state) 1 ms to 100 ms
T
10
EOM pulse width (RUN state)
50Hzpowerfrequency
T
I
=0.02PLC to 1PLC: 32.8 ms
T
I
=10PLC, 100PLC: 164 ms
T
I
=ms setting INT{(T
I
+39)×0.025}×32.8
60Hzpowerfrequency
T
I
=0.02PLC to 1PLC: 29.4 ms
T
I
=10PLC, 100PLC: 147 ms
T
I
=ms setting INT{(T
I
+39)×0.025}×29.4
T
I
: Integration time
INT (value): Rounds off the decimal portion of the value.