Conguring External Input and Output
TRIGsignaloperationwhiletheinputlterissettoON
(Chatter)
Response
time
During measurement
Hold the input signal until the response time has passed.
EOMsignaloutputtype
You can choose to either hold the EOM signal’s output until the next trigger is received or output
the set pulse.
(Measurement screen) MENU >
EXTI/O
1
2
Selecttheoutputtype.
HOLD Holds the EOM signal after
measurement is complete
(default setting).
PULSE Outputs a pulse with the set
width after measurement is
complete.
3
(After selecting PULSE)
Set the pulse width and touch ENTER.
˄ Increases the value by 1.
˅ Decreases the value by 1.
CLR Deletes all characters.
CANCEL Cancels the setting and returns to
the previous screen.
Valid setting range: 1 ms to 100 ms
(Default setting: 5 ms)
Measurement
processing
External Control (EXT I/O)
11