Comparator Measurement (Using a Single Judgment Standard)
Toperformjudgmentaftermeasuredvaluesstabilize
Since measured values may exhibit instability immediately after the instrument is connected to the
measurement target when performing measurement manually, readings may temporarily exceed
the judgment range.
When judgment delay is enabled, the judgment will be output after the same judgment is obtained
the set number of times.
(Measurement screen) MENU > MATH > COMP
1
2
Enablethejudgmentdelay.
(Default setting: OFF)
3
Selectthedesirednumberofjudgment
delays.
(Default setting: 2)
Tooutputjudgmentresultstoanexternaldeviceortoprintjudgment
results
Set the comparator function to ON, congure external output (p. 111) or printing (p. 143), and
prepare the associated equipment.