162
Functional Specications
Measured value format Operation Changes the format of responses to measured value queries.
RANGE FIX setting: Exponential part xed based on the
measurement range
FLOAT setting: Floating point
(When using the FLOAT setting, the instrument automatically
transitions to the STOP state when transitioning to the
REMOTE state.)
Setting Measured value format: RANGE FIX, FLOAT
Default setting Measured value format: RANGE FIX
SCPI ID Operation Sets the response string for the
*IDN?
query.
Setting SCPI ID: Up to 127 characters
Default setting Blank (HIOKI, model name, serial number, software version)
Remote When communicating with the LAN, USB, RS-232C, or GP-IB interface, places the
instrument in the remote state and disables touch panel and key operations.
The
[TRIG]
key functions except while the instrument is in the RUN state.
The remote state can be canceled as follows:
By pressing the LOCAL key on the touch panel
By cycling the instrument’s power
By sending the :
SYSTem:LOCal
command with the LAN, USB, RS-232C, or GP-IB
interface
By sending the
GTL
command with the GP-IB interface
Startup settings Operation Selects which settings to apply when the instrument is turned
on.
Settings Startup setting: LAST, FACTORY, PANEL
Panel: No. 01 to No. 30
Default settings Startup setting: LAST
Panel: No. 01
Reset Reset Reverts all settings other than panel data and interface
settings to their factory defaults.
(Operation is the same as for the
*RST
,
:SYSTem:PRESet
,
:STATus:PRESet
commands.)
System reset Reverts all settings to their factory defaults.