Q&A (Frequently Asked Questions)
No. Issue Itemstocheck Possiblecauses→Solutions See
2-6 Measured
temperature is not
displayed correctly.
Temperature sensor There is a problem with the
connection.
→ Connect the temperature sensor
by inserting the plug all the way in.
The specied temperature sensor
has not been used.
→ Use the Z2001 Temperature
Sensor.
p. 27
The temperature sensor is
defective.
→ Replace the Z2001 Temperature
Sensor.
-
Sub-display Trend and settings
screens
Temperature cannot be displayed
on the trend display or settings
screens.
→ Close the trend display or
settings screen.
p. 16
2-6 Measured
temperature is not
displayed correctly.
STOP or EXTERNAL trigger Temperature is updated based on
the voltage. Temperature will not
be updated when measurement is
stopped.
→ Push [TRIG] key to execute
triggered measurements or restart
continuous measurements.
p. 37
p. 38
3. Communications
Operation can be easily conrmed by referring to “8.3 Communications Settings” (p. 109).
No. Issue Itemstocheck Possiblecauses→Solutions See
3-1 The instrument is
not responding at
all.
Display REMOTE icon is not
displayed
Connection is not established.
→ Check whether the connector
has been connected.
-
→ Check whether the interface
setting is correct.
p. 98
p. 100
p. 102
p. 104
(USB)→ Install the driver in the
control device.
p. 98
(RS-232C)→ Use a cross cable. p. 101
(USB, RS-232C)→ Check the COM
port number on the control device.
p. 98
p. 100
(RS-232C)→ Use the same
communication speed for the
instrument and the control device.
p. 100
REMOTE icon is
displayed
Commands are not accepted.
→ Check the software delimiter.
-
→ (GP-IB) Check the message
terminator setting.
→ (GP-IB) Check whether
the address setting has been
congured properly.
p. 102