179
Error Displays
Display Description Solution
INFO:081 Enter password for
AdjustmentMode.
Enter the password for
adjustment mode.
-
Err.Cal Compensation values for
self-calibration are not
correct. There is a failure in
communicating with the A/
D converter due to external
noise or the instrument is
malfunctioning.
If this error is displayed
continuously, request for
service.
Err.AD A communication error with A/
D converter. There is a failure
in communicating with the A/
D converter due to external
noise or the instrument is
malfunctioning.
If this error is displayed
continuously, request for
service.
Err.REF Reference voltage error. If this error is displayed
continuously, request for
service.
Maintenance and Service
14
w ww . . co m
information@itm.com1.800.561.8187