79
Correcting Measured Values
Adjustingthezero-pointusingauser-setvalue
(Measurement screen) MENU >
MATH
> NULL
1
EnabletheNULLfunction.
(Default setting: OFF)
2
1
2
Enter a NULL value and touch ENTER.
+/− Toggles between positive and
negative values.
CLR Deletes the entire value.
BS Deletes the previous character.
CANCEL Cancels the setting and returns to
the previous screen.
(Default setting: 0 V; valid setting range:
-1000 V to 1000 V)
The zero-point will be adjusted.
The NULL icon will be displayed on the
Measurement screen.
Pressing the
[NULL]
key while the NULL function is in the ON state (while the NULL icon is
being displayed) will disable the NULL function.