Appendix 4 Temperature Compensation Table
166
Table No.1
Source
GB50150-91 Standard for hand-over test of electric equip-
ment, electric equipment installation engineering (Chinese)
Object under test
Motor stator winding: thermoplas-
tic insulating material
Selectable reference
temperature range
5 to 75°C, (default 20°C)
Convertible range of
actual temperature
used for measurement
5.0 to 70.0°C
Compensation
formula
Converted to a resistance value at
the reference temperature using
the formula below and result dis-
played.
Rtref : Resistance after compen-
sation for reference tem-
perature of tref°C
Rt : Resistance measured at
the temperature of t°C
tref : Reference temperature [°C]
t : Actual temperature used
for measurement [°C]