3.2 Measuring Insulation Resistance
1
2
3
4
5
6
7
8
9
10
11
付録
索引
69
When the display reflects the following
state, insulation resistance measurement
cannot be started.
• The setting value is blinking to indicate
that the instrument being set up
• The mark is blinking
•While [TC] is lit, the actual measurement
temperature is shown as [- - -]
• An error massage is displayed
If the indication is unstable, the average of
the measurement is shown.
Pressing the key toggles [AVE] on/
off.
While [AVE] is on, display update interval is
four seconds, normally.
But in the following case, the interval is one
second even if [AVE] is on.
• During 15 seconds after the measure-
ment started
• During 5 to 10 seconds after the mea-
surement range changed
The state not to be started the measure-
ment
Average function