4.3 Temperature Compensation (TC)
1
2
3
4
5
6
7
8
9
10
11
付録
索引
95
6. Press the key to confirm the ref-
erence temperature.
[TC] lights up and the instrument enters
temperature compensation mode.
The LCD displays the resistance at the
reference temperature converted from
the measurement.
The bar graph shows the value before com-
pensation.
Bar graph
(Value before
compensation)
Digital value
(Value after
compensation)
Actual Measurement Temperature
Reference Temperature
If the resistance before compensation
exceeds the measurement range, it
cannot be converted and the LCD
displays [
- - -
].
After the instrument is placed in
temperature compensation mode,
measurement or input of temperature and
measurement of insulation resistance
may be conducted.
However, if the instrument is placed in
temperature compensation mode when
the temperature is not held (
is off), measure or enter temperature
before measuring resistance. You cannot
measure resistance first.