116
Troubleshooting
Problem Possible cause Action
Reference
The measured value
diers for each
measurement, even
though the object
under measurement is
the same.
Material of the object under
measurement is having an
eect.
After a measurement,
wait a sucient amount of
time (1 hour to 1 day as a
general rule) before taking
a new measurement.
The eect of polarization*
2
will be more apparent for
higher insulation resistance
values.
–
Temperature and humidity
characteristics of the object
under measurement are
having an eect.
Measure in an environment
with the same temperature
and humidity. An insulating
material typically has
characteristics in which
insulation resistance
decreases as temperature
and humidity increase.
Reference: There are
certain insulated cables
in which the insulation
resistance decreases
to 1/4 or less when the
temperature increases
10
°
C.
–
*1. Even with new alkaline batteries, it may not be possible to use batteries with a
large internal resistance as the amount of obtainable energy will be low. If the
instrument does not work when using new batteries, use batteries from a dierent
manufacturer.
*2. Polarization: The phenomenon in which positive and negative charges on the
atoms of a material move in opposite directions causing a shift of the
center when an electric eld is applied to the material.
If the cause is unknown, try resetting the system.