23
Part Names and Functions
1111
Current value
Static
capacitance value
Alternates between the current value and static capacitance value.
(Automatically switches every two seconds.)
1212
Unit
Displays the unit that corresponds to the measured value.
1313
HOLD
Displayed when the measured value is retained.
1414
Indicates the comparator is enabled.
APS
Indicates the automatic power save function enabled.
1515
Indicates the remaining battery level.
1616
PV
Displayed during a PV insulation resistance measurement. (IR5051
only)
TIMER
Displayed during a TIMER measurement.
PI
Displayed during a PI (polarization index) measurement.
DAR
Displayed during a DAR (dielectric absorption ratio) measurement.
SV
Displayed during an SV (step voltage) measurement.
Ramp
Displayed during a Ramp (ramp voltage) measurement.
DD
Displayed during a DD (dielectric discharge) measurement.