Checking Recorded Data
The recording method of the displayed record is identied as follows.
Memory number is [Lr] Logging recording data
Memory number is not [Lr] Manual recording data
The type of manual record is identied as follows.
[PI] is lit PI (polarization index) data
[DAR] is lit DAR (dielectric absorption ratio) data
[SV] is lit SV (step voltage test) data
[Ramp] is lit Ramp (ramp voltage test) data
[DD] is lit DD (dielectric discharge) data
For logging records, only the last data is displayed.
For data that is not displayed on the LCD, you can display this data with the same
key operations as for switching the display method during a test.