97
Input Specications, Output Specications, and Measurement Specications
Operation
uncertainty (B)*
1
±
30% rdg
Guaranteed range
of operation
uncertainty
Greater than or equal to 0.1 M
Ω
but less than or equal to the
resistance value calculated by dividing the test voltage (setting) by
100 nA
E󰀨ect of supplied
voltage (E2) *
1
Accuracy
×
0.5 and within accuracy specications
E󰀨ect of temperature
(E3) *
1
Accuracy
×
1.0 is added to the accuracy (applicable in the ambient
temperature range excluding 18
°
C to 28
°
C)
When the L9850-11 and L9850-12 test leads (10 m) are used, the
accuracy of 501 G
Ω
or more is not guaranteed.
E󰀨ect of capacitance
component
Within
±
10% at a capacitance of 5 µF or less (including drift)
Number of possible
measurements
200 or more
Overload protection 1100 V AC (for 10 sec. between + & − terminals)
6000 V DC (for 10 sec. between + & − terminals)
Display update Once/sec. or more (when the lter is on, 0.25 times/sec.)
Response time Measurement condition: Resistive load
(output voltage 5000 V, resistive load 1 T
Ω
)
Filter o󰀨: Within 3 sec.
Filter on: Within 10 sec.
Induction noise
removal
3 mA max
*1. B, E2, and E3 are the symbols specied by IEC61557.
HIOKI IR5050A961-00