8.1 External Input/Output Connector and Signals
96
(4) Signal correspondence chart
LOAD0 to LOAD3
RNG_OUT0 to RNG_OUT3
BCDm-0 to BCDm-3
LOAD3 LOAD2 LOAD1 LOAD0 Panel number
OFF OFF OFF OFF No change
OFF OFF OFF
ON 1
OFF OFF
ON OFF 2
OFF OFF
ON ON 3
OFF
ON OFF OFF 4
OFF
ON OFF ON 5
OFF
ON ON OFF 6
OFF
ON ON ON 7
ON OFF OFF OFF 8
ON OFF OFF ON 9
ON OFF ON OFF 10
ON OFF ON ON No change
ON ON OFF OFF No change
ON ON OFF ON No change
ON ON ON OFF No change
ON ON ON ON No change
RNG_OUT3 RNG_OUT2 RNG_OUT1 RNG_OUT0 Range
OFF OFF OFF
ON 30 m
OFF OFF
ON OFF 300 m
OFF OFF
ON ON 3
OFF
ON OFF OFF 30
OFF
ON OFF ON 300
OFF
ON ON OFF 3 k
OFF
ON ON ON 30 k
ON OFF OFF OFF 300 k
ON OFF OFF ON 3 M
BCDm-3 BCDm-2 BCDm-1 BCDm-0 Measured value
OFF OFF OFF OFF 0
OFF OFF OFF
ON 1
OFF OFF
ON OFF 2
OFF OFF
ON ON 3
OFF
ON OFF OFF 4
OFF
ON OFF ON 5
OFF
ON ON OFF 6
OFF
ON ON ON 7
ON OFF OFF OFF 8
ON OFF OFF ON 9

Table of Contents