5.1 Judging Measured Values (Comparator Function)
60
The comparator judgment mode can be set as one of the following:
Before Using the Comparator Function
• The comparator judgment indicator will function as follows for over-range events
(“
OvrRng” display) and measurement faults (“- - - - -” display):
See: "Confirming Measurement Faults"(p.38)
• If power is turned off during comparator setting, changes to settings are lost as they
revert to their previous values. To accept the settings, press the .
5.1 Judging Measured Values
(Comparator Function)
Select the REF% (relative values) judgment mode
Decide whether a measured val-
ue is within specified tolerance
limits relative to a specified ref-
erence value (p. 64)
Select the ABS (absolute values) judgment mode
Decide whether a measured val-
ue is between specified upper
and lower threshold values (ab-
solute values) (p. 62)
12.000 k... reference value
±0.08%....... positive/negative
tolerance
example
100.00 m . upper threshold
80.00 m ... lower threshold
example
Hi
IN
IN
Lo
Hi
IN
Lo
Positive
tolerance [%]
Reference value []
Negative
tolerance [%]
Upper
threshold []
Lower
threshold []
Measured value display Comparator Judgment Indicator
+OvrRng Hi
-OvrRng Lo
- - - - -
Off (no judgment)