12.1 Troubleshooting
169
12
The following messages are displayed when the instrument detects an error or abnormal
measurement setting. If repair is necessary, contact your authorized Hioki distributor or
reseller.
If damage is suspected, check the "Q&A (Frequently Asked Questions)" (p.160) section
before contact your authorized Hioki distributor or reseller.
If an error is shown on the LCD and the instrument needs to be repaired, please contact
your authorized Hioki distributor or reseller.
Error Displays and Remedies
Display Description Remedy
+OvrRng/-OvrRng
Over-range (p. 38) Select the appropriate range.
ERR:001
LOW limit is higher than
UPP limit.
Cannot set because the lower
limit value is larger than the
upper limit value.
Set an upper limit value that is
larger than the lower limit value.
(p.62)
ERR:002 REF setting is zero.
Cannot set because the refer-
ence value setting is zero.
Set a reference value that is
larger than zero. (p.64)
ERR:003
Cannot enable while
comparator is ON.
Cannot switch ranges when
the comparator is ON.
Set the range after turning the
comparator OFF or select the
range to use on the Comparator
Settings screen. (p.60)
ERR:004
Cannot enable while
comparator is ON.
Cannot turn auto-ranging ON
while the comparator is ON.
Use with the comparator set to
OFF.(p.61)
ERR:010
0 ADJ error.
Must not exceed 50% f.s.
Out of zero-adjust range. The
reading must be within 50% of
range full-scale.
Check the zero-adjustment pro-
cedure (p. 44).
ERR:011
Temp. sensor error.
Cannot calculate.
Cannot perform calculations
due to a temperature sensor
error.
Check the temperature sensor.
ERR:030 Command error. Command Error.
Check for incorrect commands
(Included application disk).
ERR:031
Execution error.
(Parameter error)
Execution Error. The parame-
ter value is out of range.
Check whether the parameter
range is correct.
ERR:032 Execution error. Execution Error.
Check whether any command
has resulted in execution error
conditions.
ERR:090 ROM check sum error.
Program ROM checksum
error
The instrument is malfunction-
ing. Request repairs.
ERR:091 RAM error. CPU RAM error
The instrument is malfunction-
ing. Request repairs.
ERR:092
Memory access failed.
Main power off,
restart after 10s.
A communications error
occurred while attempting to
access the memory.
Turn off the main power switch,
wait at least 10 seconds, and
turn it back on.
ERR:093 Memory read/write error. Memory read/write test error
The instrument is malfunction-
ing. Request repairs.
ERR:095 Adjustment data error. Adjustment data error
The instrument is malfunction-
ing. Request repairs.