12.1 Troubleshooting
164
2-5 It is impossible to
perform zero-
adjustment.
Measured values before zero-adjust-
ment exceed -3% to 50% of each range
full-scale, or a measurement fault has
occurred.
There is a problem with the wiring.
Repeat zero-adjustment with
the correct wiring. Since zero-
adjustment cannot be per-
formed if the resistance value is
too high, for example with a
user-made cable, work to mini-
mize the wiring resistance.
p.A8
2-6 The auto-hold
function is not
working (hold
operation is not
being canceled).
Measured values Are unstable. See No. 2-1 above, “Measured
values are unstable.”
p.162
Do not change. An appropriate range has not
been selected.
Select an appropriate range or
use auto-ranging.
p.32
2-7 Measured tem-
perature is dis-
played
incorrectly.
The temperature sensor or ther-
mometer is not properly con-
nected.
Connect the temperature sen-
sor by inserting the plug all the
way.
The settings have been improp-
erly configured.
Check the settings.
A temperature sensor other than
that specified is used.
Model 9451 Temperature Probe
is not supported.
p.27