FAXITRON PATH USER MANUAL | HOLOGIC | REVISION 024
ECTION
Installation
nd Setup
2
2-5
Environmental Conditions
Operational:
Temperature: 15.5 to 30 degrees C or 60 to 86 degrees F.
Relative humidity: < 75 %.
Storage and transport:
Temperature: 0 to 45 degrees C or 32 to 113 degrees F.
Relative humidity: < 75 %.
This equipment is designed for the following conditions:
Indoor use
Altitude up to 2000 meters
Pollution Degree 2
Over-voltage Category II
The Faxitron Path should not be moved once installed. The system was designed as a
stationary unit to be set up on a counter or tabletop.
RADIATION SAFETY
Radiation shielded cabinet.
Compartment door equipped with dual safety interlocks.
Radiation: Less than 0.1 mR/hr at 5 cm (2 in.) from exterior surface at maximum kV.
Certified to comply with standards set by the US Food and Drug Administration, Center
for Devices and Radiological Health, (21 CFR 1020.40).
Designed to comply with standards set forth by Underwriters Laboratory the US and
Canada
Faxitron recommends that users receive training in basic ionizing radiation safety before
using the system.
EXPOSURE CONTROL: Automatic or Manual.
X-RAY DUTY CYCLE: 50%
LASER INFORMATION: Class 1 Laser Product
IEC 60825-1 Ed. 2
Complies with 21 CFR 1040.10 and 1040.11 except for
deviations pursuant to Laser Notice No. 50, dated June 24,
2007.
POWER REQUIREMENTS: 100 – 250 VAC, 50/60 Hz, 200 VA Max
HEAT OUTPUT: 500 BTU/hr Max
2.3.0
Specifications