50
12.2.1. Configuration mode operation table
Configuration mode allows the user to perform calibration and configure parameters such as full-scale range, calibration gas level, calibration interval,
inhibit current & timeout, alarm settings, relay settings, set a password, etc. To activate Configuration mode, hold the magnet over the ENTER switch
for at least 3 seconds and then remove. Configuration mode can be password protected to prevent unauthorized personnel from changing parameters.
Initially the password is set to ‘0000’ meaning it is disabled. While in Configuration mode, the output current of the detector is inhibited to prevent false
alarms.
Use the table below to help navigate the menus and make configuration changes. The menus are shown in the left-hand column. Use

to select the required menu and ‘
’ to enter. Follow the information and instructions in the table from left to right from the required
menu.
OK
OK
OK
OK
SEt
CAL
1
GAS NAME, ZERO CONCENTRATION AND
FLASHING ‘ ’ ICON DISPLAYED. APPLY ZERO
GAS AND USE ‘ ’ WHEN READING IS
STABLE. ‘ZEro PASS’ DISPLAYED IF OK, ‘ZEro
FAIL’ IF NOT
‘SPAN’ DISPLAYED AND ‘YES’
FLASHING TO ASK IF YOU NOW
WANT TO PERFORM SPAN
CALIBRATION. USE ‘ ’ TO PROCEED
OR USE TO SELECT ‘No’ AND
RETURN TO MENU MODE.
‘GAS’ TARGET CONCENTRATION
FLASHING AND ‘ ’ ICON DISPLAYED.
USE TO CHANGE GAS
CONCENTRATION AND ‘ ’ TO START
SPAN CALIBRATION.
GAS NAME, CURRENT CONCENTRATION AND
FLASHING ‘ ’ ICON DISPLAYED. APPLY SPAN GAS
AND USE ‘ ’ WHEN READING STABLE. IF OK ‘PASS’
& ‘PurG’ DISPLAYED (IF FAIL ‘SPAN FAIL’ DISPLAYED
AND RETURNS TO MENU). REMOVE SPAN GAS.
WHEN READING <50% OF SPAN POINT,
COUNTDOWN BEGINS & UNIT RETURNS TO MENU
MODE.
InIt
2
bASE
‘InIt bASE” DISPLAYED AND RETURN TO
MENU MODE WHEN COMPLETED
SEL
2
SENS
‘SEnS’ DISPLAYED WITH SENSOR TYPE
DISPLAYED. USE TO SELECT
DIFFERENT SESNOR.
IF SENSOR TYPE IS CHANGED, ‘CAL
YES’ DISPLAYED TO ASK
IF YOU NOW WANT TO PERFORM
CALIBRATION. USE ‘ ’ TO PROCEED
OR USE TO SELECT ‘No’ AND
RETURN TO MENU MODE.
‘9AS’ DISPLAYED WITH GAS TYPE
DISPLAYED. USE TO SELECT
DIFFERENT GAS.
IF GAS TYPE IS CHANGED,
‘CAL YES’ DISPLAYED TO ASK IF YOU
NOW WANT TO PERFORM
CALIBRATION. USE ‘ ’ TO
PROCEED OR USE TO SELECT
‘No’ AND RETURN TO MENU MODE.
SEL
9AS
2
SEt
rAn9
BAR GRAPH INDICATING CURRENT RANGE,
‘rAn9’ DISPLAYED & CURRENT RANGE
FLASHES. USE TO SELECT DIFFERENT
RANGE
IF RANGE IS CHANGED, ‘CAL’
DISPLAYED AND ‘YES’ FLASHING TO
ASK IF YOU NOW WANT TO
PERFORM CALIBRATION. USE ‘ ’ TO
PROCEED OR USE TO SELECT
‘No’.
IF RANGE IS CHANGED, ‘ALrm’
DISPLAYED AND ‘YES’ FLASHING TO
ASK IF YOU NOW WANT TO MODIFY
ALARM SETTINGS. USE ‘ ’ TO
PROCEED OR USE TO SELECT
‘No’ AND RETURN TO MENU MODE.
UNIT RETURNS TO MENU MODE
‘Inhb’ DISPLAYED WITH ‘ ICON
‘tImE’ DISPLAYED WITH ‘ ICON.
FLASHING CURRENT INHIBIT
TIMEOUT PERIOD (MINUTES)
FLASHES. USE TO SET NEW
TIMEOUT. (IF SET TO 0 OUTPUT IS
PERMANENTLY INHIBITED)
ConF
Inhb
4
FLASHING. CURRENT INHIBIT mA VALUE
FLASHES. USE TO SELECT NEW VALUE
(2 OR 4mA TOXICS, OR 2 OR 17.4mA
UNIT RETURNS TO MENU MODE
OXYGEN VERSION).
SEt
PASS
‘PASS’ DISPLAYED WITH FIRST DIGIT PLACE
ICON OF THE PASSCODE FLASHING.
USE TO SELECT 1
ST
DIGIT OF CURRENT
PASSWORD. USE TO MOVE TO NEXT
DIGIT AND SET REST OF PASSCODE.
UNIT RETURNS TO MENU MODE
Sensepoint XCD Technical Manual SPXCDHMANEN Issue 15