Sensepoint XCD Technical Manual SPXCDHMANEN Issue 15
7
Special conditions for Sensepoint XCD with Oxygen
Sensor
1. Compliance with the Technical Manual is required for safe use of Sensepoint XCD.
2. The following environmental conditions of operation apply for the use of Sensepoint XCD
with oxygen sensor:
Temperature: -20°C to +55°C
Relative humidity: 10 % to 90 %
Pressure: 80 kPa to 120 kPa
(extended range for temperature according to EN 50104)
3. In order to minimise measurement errors, the ambient conditions (temperature, pressure,
humidity) during the calibration of the measuring channels should be as close as possible
to those to be encountered during normal operation.
4. Measuring values from 20.7 % v/v up to 21.1 % v/v are indicated as
20.9 % v/v oxygen during operation in the measuring mode. Measuring values below 3
% v/v oxygen are indicated as 0 % v/v oxygen (analogue output 4 mA).
5. Measuring values above the end of the measuring range are indicated as 25 % v/v O
2
and
blinking W04. The analogue output is set to 22 mA.
Fault messages are shown on the display as F-XX and the analogue output is set to
1 mA.
6. Special states for Analogue output:
≤ 1 mA: System and other faults
2 mA: Start up, Inhibit
22 mA: Over ranging
7. If the Sensepoint XCD is operating in conjunction with a control unit to analyse the
measuring signal, the control unit must operate correctly when the XCD transmits a value
outside of the measuring range.
8. Use without Weather Protection Cup accessory (P/N SPXCDWP), only.
9. If substances (e. g. sensor poisons) that could interfere with and affect the sensitivity of
the sensing device are to be expected in the atmosphere to be monitored which may cause
a rapid change of sensitivity, the calibration interval shall be reduced.
10. Before use, ensure that any alarm delay set is suitable for the intended application.
11. Before use, consideration must be given to cross sensitivity to other gases. Refer to the
Sensepoint XCD Technical Manual and the Sensor Datasheet for cross interference data.
12. Some types and concentration of dust in the measured atmosphere may impair the
measuring function of the gas detector.
13. The Type Examination Certificate applies to the measurement of oxygen up to 25 % v/v.

Other manuals for Honeywell Sensepoint XCD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Honeywell Sensepoint XCD and is the answer not in the manual?

Honeywell Sensepoint XCD Specifications

General IconGeneral
TypeFixed Gas Detector
Ingress ProtectionIP66/67
DisplayLCD
Output4-20 mA, Relay
CertificationATEX, IECEx
Sensor TypeElectrochemical, Infrared
Humidity0 to 99% RH (non-condensing)
MaterialStainless Steel
Detection PrincipleDepends on Sensor Type
Power Supply24 VDC nominal (17-32 VDC)