1/99 SMV 3000 Transmitter Users Manual 131
11.4 Diagnostic Messages, Continued
Diagnostic Messages,
continued
Table 32 Non-Critical Status Diagnostic Message Table, Continued
SMV Status SCT Status Message SFC Display Message TDC Status Message Possible Cause What to Do
9-7 Reynolds Number is Out
of Range
- STATUS 9-7
The high or low Reynolds number
limit was exceeded.
Verify high or low Reynolds
number limit.
Calculate Reynolds number
for flow conditions causing the
message.
8-7 Sensor Mismatch PV3
SAVE/RESTORE
TYPE MISMATCH
SNSR MISMTCH PV3
Number of wires selected does not
match number of sensor wires
physically connected to the
transmitter.
Check sensor wiring and type.