4.
Set the
HP
5334B
as follows:
COM A
...................................................................................................................
OFF
CHAN A and
B 500
............................................................................................
OFF
Observe:
The
HP
5334B
front panel displays
10.000 000 0
MHz
+
0.2
Hz.
Channel A trigger light flashing
and Channel
B
light not flashing.
5.
Set the
HP
5334B
as follows:
CHAN A
50i2
........................................................................................................
ON
Observe:
The
HP
5334B
front panel displays
all
dashes and Channel A trigger light stops flashing.
6.
Set:
........................................................................................................
CHAN
A
50 R
OFF
Observe:
Condition prior to switching in
50R
impedance.
7.
Set:
CHAN A
XI0
A'ITN
...........................................................................................
ON
Observe:
The
HP
5334B
front panel displays all dashes and Channel A trigger light stops flashing.
8.
Set:
...........................................................................................
CHAN A
XI0
ATTN OFF
Observe:
Condition prior to switching in
XI0
attenuator.
9.
Set:
100
kHz
FILTER A
..............................................................................................
ON
Observe:
The
HP
5334B
front panel displays all dashes and Channel A trigger light stops flashing.
10.
Set:
..............................................................................................
100
kHz
FILTER A OFF
Observe:
Condition prior to switching in
100
kHz
Filter.
11.
Set:
...................................................................................................................
COM A ON
...........................................................................................................
FUNCTION FREQ
B
Connect the rear panel TIME BASE oscillator signal to the Channel
B
Input.
Observe:
The
HP
5334B
front panel stops updating and the trigger lights stop flashing.
12.
Set:
...................................................................................................................
COM A OFF
Observe:
The
HP
5334B
front panel displays
10.000 000 0
MHz
2
0.2
Hz.
Channel
B
trigger light flashing
and Channel A trigger light not flashing.
13.
Set:
CHAN
B 50R
........................................................................................................
ON
HP
5334B
-
Service Manual
4-7