EasyManua.ls Logo

HP 5334B - 4-21. T.I. A+B

HP 5334B
242 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Observe:
The
HP
5334B front panel displays all dashes and Channel
B
trigger light stops flashing.
14.
Set:
CHAN
B
500
........................................................................................................
OFF
I
Observe:
Condition prior to switching in 50n impedance.
15. Set:
CHAN B
XI0 ATTN
...........................................................................................
ON
Observe:
The
HP
5334B front panel displays all dashes and Channel
B
trigger light stops flashing.
16.
Set:
CHAN
B
XI0 ATTN
...........................................................................................
OFF
Observe:
Condition prior to switching in XI0 A'ITN.
Test Record:
Mark Pass or Fail on the Operational Verification Test Record Card, line
5.
What
Checked:
Relays and circuitry.
For Failures:
Refer to Section VIII of this manual. The Input Amplifier and Executive Blocks are the main com-
ponents of this test. Other blocks involved are DAC, Measurement, and Front Panel blocks.
4-21.
T.I.
A+B
Description:
Slope switch verification.
Procedure:
1.
Connect the rear panel TIME BASE oscillator signal to the Channel A Input.
2.
Set the
HP
5334B as follows:
Reinitialize the 5334B.
C0M.A
................................................................................................................
ON
TRIGGER LEVEL controls
..............................................................................
set to OV
+
0.2V
using READ LEVELS
"L"
mode
AUTO TRIG
.........................................................................................................
OFF
CHAN A and B 500
............................................................................................
ON
GATE TIME
.........................................................................................................
1
Second
FUNCTION
...........................................................................................................
T.I. A+ B
Observe:
The HP 5334B front panel displays
0
ns
+-
6
ns.
3.
Set both:
Channel A and B Negative SLOPE
..................................................................
ON
(Counter now triggers on negative slope.)
Observe:
The HP 5334B front panel displays 0 ns
+-
6
ns.
HP
5334B
-
Service Manual
4-8