Description:
The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
four different frequency settings.
Frequencies and conditions tested:
10
Hz
and
20
MHz,
dc coupled,
1
Mfl
30
Hz
and
20
MHz,
ac coupled,
1
Mfl
1
MHz
and
20
MHz,
ac coupled,
50fl
HP5334B
UNIVERSAL COUNTER
I
REF
OUT
'
I
5Oll
FEE
DTHROUGH
HP3325A
SYNTHESIZER/
FUNCTION
GENERATOR
1
REF
IN
.................
00000
q
q
=..
==
ooooo
110
0
DO000
To@
-
-
--
Figure
4-2.
Channel
A
Frequency and Sensitivity Setup,
10
HZ
-
20
MHz
Equipment:
..............................................................................................
Function Generator
HP
3325A
Procedure:
1.
Set the function generator as follows:
Frequency
...............................................................................................................
10
Hz
Amplitude
..............................................................................................................
15
mV
rms
.................................................................................................................
Function Sine Wave
2.
Set the
HP
5334B as follows:
Reinitialize the
5334B.
FUNCTION ........................................................................................................... FREQ
A
SENS
................................................................
......................................................
ON
CHAN
A
TRIGISENS control
...........................................................................
fully
CW
HP
5334B
-
Service
Manual
4-14