HP 5334B - 4-28. CHANNEL A FREQUENCY RESPONSE AND SENSITIVITY TEST, 10 Hz-20 MHz

HP 5334B
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Description:
The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
four different frequency settings.
Frequencies and conditions tested:
10
Hz
and
20
MHz,
dc coupled,
1
Mfl
30
Hz
and
20
MHz,
ac coupled,
1
Mfl
1
MHz
and
20
MHz,
ac coupled,
50fl
HP5334B
UNIVERSAL COUNTER
I
REF
OUT
'
I
5Oll
FEE
DTHROUGH
HP3325A
SYNTHESIZER/
FUNCTION
GENERATOR
1
REF
IN
.................
00000
q
q
=..
==
ooooo
110
0
DO000
To@
-
-
--
Figure
4-2.
Channel
A
Frequency and Sensitivity Setup,
10
HZ
-
20
MHz
Equipment:
..............................................................................................
Function Generator
HP
3325A
Procedure:
1.
Set the function generator as follows:
Frequency
...............................................................................................................
10
Hz
Amplitude
..............................................................................................................
15
mV
rms
.................................................................................................................
Function Sine Wave
2.
Set the
HP
5334B as follows:
Reinitialize the
5334B.
FUNCTION ........................................................................................................... FREQ
A
SENS
................................................................
......................................................
ON
CHAN
A
TRIGISENS control
...........................................................................
fully
CW
HP
5334B
-
Service
Manual
4-14