3.
Connect the function generator signal to the HP
5334B
Input
A
using a
50n
feedthrough connector as
shown
in
Figure
4-2.
VERIFY:
The Counter displays
10
Hz
+
0.03
Hz.
4.
Record the measurement on the Performance Test Record Card, line
1.
5.
Set the function generator as follows:
Frequency
...............................................................................................................
20
MHz
VERIFY:
The Counter displays
20
MHz
+
0.3
Hz.
6.
Record the measurement on the Performance Test Record Card, line
2.
7.
Set the function generator as follows:
Frequency
...............................................................................................................
30
Hz
8.
Set the HP
5334B
as follows:
...........................................................................................................................
AC
ON
VERIFY.
The Counter displays
30
Hz
+
0.03
Hz.
9.
Record the measurement on the Performance Test Record Card, line
3.
10.
Set the function generator as follows:
...............................................................................................................
Frequency
20
MHz
VERIFY:
The Counter displays
20
MHz
+
0.3
Hz
11.
Record the measurement on the Performance Test Record Card, line
4.
12.
Set the function generator as follows:
Frequency
...............................................................................................................
1
MHz
NOTE
Remove the
50R
feedthrough connector.
13.
Set the HP
5334B
as follows:
..........................................................................................................................
50R
ON
VERIFY:
The Counter displays
1
MHz
*
0.04
Ilz.
14.
Record the measurement on the Performance Test Record Card, line
5.
HP
5334B
-
Scrvicc Manual
4-15