6.
Record the measurement on the Performance Test Record Card, line 14.
...............................................................................................................
7.
Set the signal generator as follows:
Frequency 80 MHz
Amplitude
..............................................................................................................
35 mV rms
8.
Set the HP
5334B as follows:
...........................................................................................................................
AC
ON
VERIFY:
The Counter displays 80 MHz
r
2 Hz.
9.
Record the measurement on the Performance Test Record Card, line 15.
10. Set the signal generator as follows:
Frequency
.............................................................................................................
100 MHz
Amplitude
..............................................................................................................
35 mV rms
VERIFY:
The Counter displays 100 MHz
+
2 Hz.
11.
Record the measurement on the Performance Test Record Card, line 16.
12. Set the signal generator as follows:
Frequency
...............................................................................................................
80 MHz
...................................
Amplitude
:
.........................................................................
35 mV rms
NOTE
Remove the
50R
feedthrough connector.
13.
Set the HP
5334B as follows:
..........................................................................................................................
50R
ON
VERIFY:
The Counter displays 80 MHz
+
2
Hz.
14. Record the measurement on the Performance Test Record Card, line 17.
15. Set the signal generator as follows:
Frequency
...............................................................................................................
100
MHz
Amplitude
..............................................................................................................
35 mV rms
VERIFY:
The Counter displays 100 MHz
+
2
Hz.
16. Record the measurement on the Performance Test Record Card, line 18.
Failure:
If any of these tests fail, refer to Section
V,
Adjustments, paragraphs 5-16 and 5-18 as a first step in
troubleshooting.