9.
Set the signal generator as follows:
Frequency
...............................................................................................................
1300
MHz
Amplitude
..............................................................................................................
+
0.5
dBm
NOTE
This amplitude provides
75
mV rms to the Channel C Input when using the
10
dB attenuator.
VERIFY:
The Counter displays 1300
MHz
+
20
Hz.
10. Record the measurement on the Performance Test Record Card, line
56.
Failure:
Perform Section
V,
Adjustments, paragraphs
5-21
and
5-22.
HP
5334B
-
Service
Manual
4-36