CHECKPOINT SUMMARY 
1  Power-up Preset 
2  Remote, Local, Local Lockout 
3  Function Select 
4  lnput Conditioning 
- 
Channel A 
5 
lnput Conditioning 
- 
Channel 
B 
6 
Trigger Levels 
7 
Gate Time 
8 
External Arming 
9 
Math Functions 
10  Memory Recall (HP 
5334A only) 
11 
Service Request 
12  Status Byte 
13  Send Error Code 
14  High Speed Output 
5. 
The next screen provides the option to receive a printed version of  this summary. 
6. 
Equipment setup instructions are provided (with reference made to the manual), then the 
HP 85 
searches 
for the address of  the 
HP 5334B. 
This search eliminates the need to set the 
HP 5334B 
to a particular 
HP-IB 
address and thus, allows an arbitrary address setting. 
If 
the address is not found, one or more of the follow- 
ing may be the cause: 
HP 5334B 
- 
not powered-up 
- 
in 
TALK 
ONLY 
mode (Addr 
50) 
Interface 
- 
not connected 
- 
defective 
Address of 
HP 5334B 
- 
set to 
721 
(controller address) 
- 
set' to 
731-749 
(not valid addresses) 
7. 
After these initial steps, the program begins the checkpoint execution. The 
HP 85 
display provides the 
description and operator instructions as each checkpoint is performed. 
8. 
At the end of  most of  the checkpoints, the 
HP 85 
displays the following prompt: 
HP 5334B 
- 
Service Manual 
4-41