Performance
tests
Introduction
The performance of the HP E1440A can be tested at three levels:
Functional Verification Tests
-
These check that the instrument
functions are working. Use these tests when you want to check that
the module is connected properly and responding to commands
from the controller. No access to the interior of the module is
necessary.
Operational Verification Tests
-
These check that the instrument
meets critical specifications. Use these tests after a repair, or as an
aid to troubleshooting. These tests can be carried out
without
a
controller.
rn
Performance Verification Tests
-
These check that the instrument
meets all warranted specifications. Use these tests as an instrument
calibration check, after repair or as an aid to troubleshooting.
Note
3
The
HP
E1440A specifications, given in Chapter 1, are valid after a
30
minute warm up period.
All test equipment used must also be allowed to warm up before
testing the HP E1440A. Refer to the equipment specifications to find
the warm up time required.
Test Record
You can record the results of the
Performance Verification Tests
using the Test Record provided at the end of this chapter. You can
reproduce this Performance Verification Test Record without written
permission from Hewlett-Packard.
Performance tests
C-1