I
DC
Offset
":VOLT:OFFS 1.499"
":VOLT:OFFS -1.499"
":VOLT:OFFS 499.9 MV"
":VOLT:OFFS -499.9 MV"
":VOLT:OFFS 149.9 MV"
":VOLT:OFFS -149.9 MV"
":VOLT:OFFS 49.99
MV"
":VOLT:OFFS -49.99 MV"
":VOLT:OFFS 14.99 MV"
":VOLT:OFFS -14.99 MV"
":VOLT:OFFS 4.999 MV"
":VOLT:OFFS -4.999 MV"
":VOLT:OFFS 1.499 MV"
":VOLT:OFFS -1.499
MV"
High-Voltage Output (Option
001)
Tolerances
This procedure continues from the previous one.
8.
Remove the 500 feedthru termination and connect the
HP
E1440A output directly to the voltmeter input.
9. Select the high-voltage output on the
HP
E1440A.
OUTPUT Vxi;":OUTP:AMPL ON"
10. Set the
HP
E1440A
DC
offset to 20
V.
OUTPUT Vxi
;
"
:VOLT
:
OFFS
20"
11.
The voltmeter reading should be +19.775 to 20.225
V.
12. Switch off the high voltage output.
OUTPUT Vxi
;
It
:
OUTP :AWL
OFF"
Performance tests