+200,CbCommand not allowed; Cannot accept command
in
mode
SWEEPJLIST"
Triggered by: all commands that attempt to alter parameters in the
SWEEPILIST subsystem. Also *SAV, *TST? and
CAL?.
Switch
FREQ:MODE to CW or FIX beforehand.
+200,"Command not allowed; Cannot accept fiequency for DC function"
Triggered by :FREQ command. Attempt has been made to
enter frequency while output function is DC, the two actions are
incompatible
+2OO,"Command not allowed; Cannot accept phase for
D
C function"
Triggered by :PHAS command. Attempt to enter phase while output
function is DC, this is not possible
+2OO,"Command not allowed; Parameter is fixed for
TTL
function"
Triggered by :VOLT or V0LT:OFFS command. The main output is
always off in TTL function, the only output is the SYNC output.
The SYNC signal characteristic is not alter able.
-213,"Init ignored; Sweep is currently running or frequency mode is
FIX"
Triggered by :INIT. Attempt to start a sweep while sweep is already
running or while frequency mode is fixed (non sweeping).
-221,"Settings codict; AC Voltage in Vrms/dBm out of range in new
function"
-
Triggered by :FUNC command. The instrument maintains the
output voltage over function changes in the current default unit. If
the current default unit is not
V
(VRMS or dBm)
a
value legal for
one output function, may not be legal for another.
In
other words
:
the Vpp voltage that represents
X
dBm in SIN
function is different from the Vpp voltage for
X
dBm in SQU
function
.
-221,"Settings confiict; Active marker in logarithmic interval"
Triggered by :FREQ:MODE command. One of the in-sequence
intervals used, has spacing LOG and a marker ON.
-221,"Settings confiict; Active marker outside sweep span"
Triggered by :FREQ:MODE command. There is a marker switched
on outside the frequency span defined
by
start and stop.
-221,"Settings codict; Already another marker trigger on"
Triggered by :OUTP:TTLT:STAT and 0UTP:TTLT:SOUR
command. Cannot accept marker trigger on, because another marker
trigger is on. The marker trigger can only be applied to one line.
-221,"Settings confiict
;
Already another sync trigger on"
Triggered by :OUTP:TTLT:STAT and 0UTP:TTLT:SOUR
command. Cannot accept sync trigger on, because another sync
trigger is on. The sync trigger can only be applied to one line.
E-2
Error
Messages