-221,"Settings codict; Start frequency
too
high
for
TTL
trigger
lines"
Triggered by: FREQ:MODE command.
If
SYNC signal is routed to
one of the
8
TTL
trigger lines and the line driver is enabled, the start
frequency is limited to 10 MHz (VXI back plane bandwidth).
-221,"Settings conflict; Stop frequency of
LOG
interval
is
lower than
start"
Triggered by: FREQ:MODE command. LOG interval has been used
with negative span (stop
<
start).
-221,"Settings codict; Stop frequency too high for post ampuer"
Triggered by: FREQ:MODE command. If post amplifier state ON,
the stop frequency is limited to
1
MHz.
-221,"Settings conflict; Stop frequency too high for
TTL
trigger lines"
Triggered by: FREQ:MODE command. If SYNC signal is routed to
one of the
8
TTL trigger lines and the line driver is enabled, the stop
frequency is limited to 10 MHz (VXI back plane bandwidth).
-221,"Settings conflict; Sync and Marker cannot drive the same trigger
line"
Triggered by: 0UTP:TTLT:STAT and 0UTP:TTLT:SOUR
command. Cannot accept sync trigger and marker trigger on the
same line
-221,"Settings codict; Time distance start-mark-stop too low
(<1.5
ms)"
I
Triggered by: FREQ:MODE command. The minimal distance
between start
->
marker, marker
->
marker or marker
->
stop is
1.5
ms.
-222,"Data out of range; AC Voltage is out of range"
Triggered by: VOLT command. Requested AC voltage is out of
absolute bounds.
-222,"Data out of range; AC Voltage is incompatible with DC offset"
Triggered by: VOLT or V0LT:OFFS command. The peak output
level (abs(0FFSET)
+
112
VOLTAGE) that would result, exceeds
the output amplifier limits.
-222,"Data out of range; Active marker is out of interval"
Triggered by: FREQ:MODE command. One of the switched ON
markers is outside the interval borders. There is a
1.5
ms minimal
distance between the marker and the start/stop point.
-222,"Data out of range; Any list
frequency
is out of range"
Triggered by: FREQ:MODE command. Any one of the list
frequencies is too high or negative.
-222,"Data out of range; A list dwell time
is
out of range"
Triggered by: 0UTP:AMPL command.
E-4
Error
Messages