-222,"Data out of range; Sweep span is out of range"
Triggered by: FREQ:MODE command. The requested sweep span
would require a startlstop frequency outside the legal range.
-222,"Data out of range; Sweep time is out of range (10 ms
..
100000 s)"
Triggered by: FREQ:MODE command. The SWE:TIME or
SWE:RTIM value is out of range.
-222,"Data out of range; Sweep time too low
(
<
10 ms
)"
Triggered by: FREQ:MODE command. The SWEEP:TIME or
SWE:RTIM or one of the dwell values used in sequence, is less than
10
ms.
-222,"Data out of range; Too many points in sweep sequence (>100)"
Triggered by: FREQ:MODE command. The expanded length of the
sweep sequence is greater than
100.
-222,"Data out of range; Trigger line index is out range"
Triggered by: 0UTP:TTLT:STAT and 0UTP:TTLT:SOUR
command. Attempt to set trigger line index out the range
0
-
7.
-240,"Hardware error; Device bus transfer timed out9'
Triggered by: asynchronous check and self test. Device bus transfer
poll routine timed out, this is a hardware fault!
-240,"Hardware error; Voltage calibration failed"
Triggered by: FUNC or CAL? command. This is a hardware fault.
I
Please report error and returned value from CAL? command to
service personnel.
-241,"Hardware missing; Opt. 001 (post ampaer) not installed"
Triggered by: 0UTP:AMPL command post amplifier option
001
is
missing. Hardware not installed.
-
330,"Self- test failed; Device bus read failed"
Triggered by: *tst? command This is a hardware fault. Please report
error and returned value from *TST? command to service personnel.
-330,"Self-test Wed; Device bus read/write failed"
Triggered by: *tst? command Device bus read or write does not
work. This is a hardware fault. Please report error and returned
value from
"TST? command to service personnel.
-330,"Self-test failed; Frac.-N chip readlwrite (pat tern:
0600000000000000) failed"
Triggered by: *tst? command This is a hardware fault. Please report
error and returned value from
"TST?
command to service personnel.
-330,"SelGtest failed; Frac.-N chip read/write (pattern:
0598765432101200) failed"
Triggered by: *tst? command This is a hardware fault. Please report
error and returned value from 'TST? command to service personnel.
-
E-6
Error
Messages