*OPC
-
*oPc
Operation complete command.
Syntax
*OX
Definition
If
"*OPC" is sent to the
HP
E1440A
while a sweep is running, then
the OPC-bit in the standard event status register is set when the
current sweep stops. If no sweep is running at the time
"OPC is sent
to the
HP
E1440A
the bit is set immediately.
The following actions cancel
*OPC
(device goes to Operation
Complete, Command Idle State):
Power-on
the
dcas
line on the interface is asserted.
rn
*CLS
*RST
Related Commands
*oPc?,
*MI
Example
OUTPUT 70911; Ii*CLS; *ESE
1
;
*SRE 32"
OUTPUT 7091
1
;
il*OPC1'
Command Reference