94 Excelon Operators Manual
Fuse Replacement in Blocker
Blocker Lamp Replacement
In case of Blocker Lamp Replacement, be sure to use the lamp described below.
Unfasten two e screws in (A) disassemble the Lamp Fixing Plate.
Remove the Lamp.
Lamp : 6V 8W (Optic Lamp)
Lamp Life :Approximately 5,000 hours (it may vary depending on the
working condition)
Be sure not to unfasten the screws above (A). Otherwise, it may cause the align
problem and malfunction.
!