Huvitz Excelon - 7.2 Automatic Calibration of Edger

Huvitz Excelon
101 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Excelon Operator’s Manual 85
7.2 Automatic Calibration of Edger
Three kinds of calibration can be done with Standard edging jig. Those are Size
Calibration, Feeler Calibration and Safety Bevel & Grooving calibration.
Location : Engineer Mode Calibration
Size Calibration
Procedures for size calibration
① Execute ‘Size Calibration’ item
② Replace Block Adapter with Standard JIG
③ Press [F6] to execute or [F7] to return to menu
④ After calibration, press [F1] to save or [F2] to discard