7.2 Automatic Calibration of Edger
Three kinds of calibration can be done with Standard edging jig. Those are Size
Calibration, Feeler Calibration and Safety Bevel & Grooving calibration.
Location : Engineer Mode Calibration
Size Calibration
Procedures for size calibration
Execute Size Calibration item
Replace Block Adapter with Standard JIG
Press [F6] to execute or [F7] to return to menu
After calibration, press [F1] to save or [F2] to discard