29
UK
10.4.5 Table repeatability and accuracy
Values for sampling rate 15 Hz, extraneous light of max. 40 klx*
Distance
in [cm]
Repeatability Accuracy
white
90 % remission
grey
18 % remission
white
90 % remission
grey
18 % remission
20�0���100�0 ± 0�45 cm ± 0�6 cm ± 1�5 cm ± 1�6 cm
100�0���200�0 ± 0�5 cm ± 0�8 cm ± 1�5 cm ± 1�8 cm
200�0���400�0 ± 1�6 cm ± 1�9 cm ± 2�5 cm ± 3�0 cm
400�0���600�0 ± 2�4 cm ± 3�3 cm ± 3�5 cm ± 4�5 cm
600�0���1000�0 ± 5�0 cm — ± 6�5 cm —
Values for sampling rate 15 Hz, extraneous light of 40...100 klx*
Distance
in [cm]
Repeatability Accuracy
white
90 % remission
grey
18 % remission
white
90 % remission
grey
18 % remission
20�0���200�0 ± 1�4 cm ± 1�4 cm ± 2�4 cm ± 2�4 cm
200�0���400�0 ± 2�5 cm ± 3�0 cm ± 3�5 cm ± 4�0 cm
400�0���600�0 ± 3�1 cm ± 4�5 cm ± 4�1 cm ± 5�5 cm
600�0���1000�0 ± 6�0 cm — ± 7�0 cm —
Values for sampling rate 1 Hz, extraneous light of max. 40 klx*
Distance
in [cm]
Repeatability Accuracy
white
90 % remission
grey
18 % remission
white
90 % remission
grey
18 % remission
20�0���100�0 ± 0�40 cm ± 0�45 cm ± 1�4 cm ± 1�5 cm
100�0���200�0 ± 0�45 cm ± 0�6 cm ± 1�5 cm ± 1�6 cm
200�0���400�0 ± 1�35 cm ± 1�4 cm ± 2�3 cm ± 2�4 cm
400�0���600�0 ± 1�9 cm ± 2�1 cm ± 2�9 cm ± 3�1 cm
600�0���1000�0 ± 3�7 cm — ± 4�7 cm —