NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
142
DIAGNOSTICS
Diagnostics offers the possibility of performing the self-test without rebooting the
NanoPhotometer
®
. Make sure that the measurement head is clean and no sample is applied
and push the self-test button to start the self-test. As result reference and signal counts and
the wavelength shift are shown. If the self-test is fails, please contact the Implen support
(support@implen.de).
REMOTE ACCESS
Option to enable remote acces, ping and Rest API Push Services.
Remote access and ping are active after booting the NanoPhotometer
®
and are
automatically switched off after 15 minutes. A renewed activation for remote access is
possible for 30 minutes at a time. The activation of ping remains active until the next restart
of the NanoPhotometer.
Rest API push service might be necessary for integration to LIMS systems. It is default off
and can be switched on if necessary. It will stay active after switched on.