NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
89
instrument version:
BCA default value 750 nm
Bradford default value 750 nm
Lowry default value 405 nm
Note: It is recommended to use the default baseline for each assay.
5. Select curve fit type: Options are linear regression, zero regression (forces the straight line
through the origin) and 2
nd
order regression
6. Select Unit
7. Add up to 20 Concentrations by pushing on the Add Concentration button. Added
concentrations can be deleted with
Enter the concentrations of the standards for the standard curve.