EasyManuals Logo

Implen NanoPhotometer NP80 User Manual

Implen NanoPhotometer NP80
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #102 background imageLoading...
Page #102 background image
NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
102
If it is desired to heat the sample to 37°C use the toggle switch to turn on cell holder heater.
When the cuvette holder has reached 37°C the toggle color changes to green.
Note: Only available for cuvette applications (NP80 and C40).
For NanoVolume application with submicroliter cell (C40 only):
Select Lid depending on your sample concentration (for concentration ranges of different
lids see page 52 Submicroliter Cell Measurements Basics).
Required sample volume for each lid is shown.
Note: A new Blank is recommended when changing the dilution lid.
4. Enter desired wavelength () to be measured. It is possible to measure up to 20
wavelengths simultaneously. More wavelength () options can be added by selecting the
Add Wavelength button. Added wavelength can be deleted with
5. Baseline correction is set off by default. Enabling the baseline corrections shows a list with
different wavelength options: 377 nm, 604 nm, 650 nm, 770 nm (N/A N50) and 823 nm
(N/A N50). Option to enter any wavelength between 200 nm and 900 nm (N50: 650 nm).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer NP80 and is the answer not in the manual?

Implen NanoPhotometer NP80 Specifications

General IconGeneral
BrandImplen
ModelNanoPhotometer NP80
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals