NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
72
Options are: 0.5 mm, 1 mm, 2 mm, 5 mm and 10 mm
If it is desired to heat the sample to 37°C use the toggle switch to turn on cell holder heater.
When the cuvette holder has reached 37°C the toggle color changes to green.
Note: Only available for cuvette applications (NP80 and C40).
For NanoVolume application with submicroliter cell (C40 only):
Select Lid depending on your sample concentration (for concentration ranges of different
lids see page 52 Submicroliter Cell Measurements Basics).
Required sample volume for each lid is shown.
Note: A new Blank is recommended when changing the dilution lid.
5. The wavelength for protein measurements can be changed in the range of
200 330nm depending on the wavelength peak of the protein. Default
setting is 280 nm
6. Select the Units in which the concentration should be calculated. Options are ng/µl, µg/µl,
µg/ml and mg/ml (default).
7. The background correction is enabled at 320 nm by default. Selection options are 320 nm,
340 nm or any wavelength in the range of 220 350 nm. The background correction can
be disabled with the toggle switch.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer NP80 and is the answer not in the manual?

Implen NanoPhotometer NP80 Specifications

General IconGeneral
BrandImplen
ModelNanoPhotometer NP80
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals