NanoPhotometer
®
N120/NP80/N60/N50/C40 User Manual Version 4.3.1
73
8. Air bubble recognition is disabled by default. When enabled it detects air
bubbles, lint residues and poor conditions of the sample.
Note: Lint residues and bad sample conditions are detected even if the
air bubble recognition is set to off.
9. For dye labelled samples, tick the dye labels in the list which should be used for result
calculation.
Note: If the used dye is not available in the list, please go to preferences and add a custom
dye to the dye list.
There is an option for dye correction which can be enabled/disabled with a toggle switch.
Note: Dye correction is only available for single dye selection.
10. Option to set/calculate a dilution factor for manual diluted samples.